  1.     SECTION 11612
  2.  
  3.     BIOLOGICAL SAFETY CABINETS
  4.  
  5. NOTE ** Labconco Corporation, biological safety cabinets.
  6. NOTE ** 
  7. NOTE ** This section is based on the products of Labconco Corporation, which is located at:
  8. NOTE **    8811 Prospect Avenue
  9. NOTE **    Kansas City, MO  64132
  10. NOTE **    Tel:  (816) 333-8811
  11. NOTE **    Tel:  (800) 821-5525
  12. NOTE **    Fax:  (816) 363-0130
  13. NOTE **    E-mail:  labconco@labconco.com
  14. NOTE **    Home Page:  www.labconco.com
  15.  
  16. NOTE ** Labconco Corporation is a manufacturer of laboratory equipment, including fume hoods, self-contained filtered laboratory enclosures, and biological safety cabinets.  This section covers part of their product line:  "Purifier" Class II, Type A and A/B3 biological safety cabinets in 3, 4, and 6 foot (0.914, 1.219, and 1.829 m) widths.  See their other sections for laboratory fume hoods and self-contained filtered laboratory enclosures.
  17. NOTE ** 
  18. NOTE ** SECTION 11612 - BIOLOGICAL SAFETY CABINETS, Copyright 1998, The Architect's Catalog, Inc.
  19.  
  20. PART 1    GENERAL
  21.  
  22. 1.1    SECTION INCLUDES
  23. NOTE ** Delete item below not required for project.  If both types of biological safety cabinet are required for project, carefully modify this section for specific applications.
  24.  
  25.     A.    Bench-mounted Class II, Type A biological safety cabinets.
  26.  
  27.     B.    Bench-mounted Class II, Type A/B3 biological safety cabinets.
  28.  
  29. 1.2    RELATED SECTIONS
  30. NOTE ** Delete any sections below not relevant to this project; add others as required.
  31.  
  32.     A.    Section 11600 - Laboratory Equipment
  33.  
  34.     B.    Section 12351 - Laboratory Casework
  35.  
  36.     C.    Section 15400 - Plumbing Fixtures and Equipment
  37.  
  38.     D.    Section 16200 - Electrical Power
  39.  
  40. 1.3    REFERENCES
  41. NOTE ** Delete references below that are not required by the text of the edited section.  The first three are for 115V/60Hz models only; the last two are for 230V/50Hz models only.  Add other references as required.
  42.  
  43.     A.    CAN/CSA - C22.2 No. 1010.1.
  44.  
  45.     B.    NSF International - Standard 49 for Biohazard Cabinetry.
  46.  
  47.     C.    UL - Standard 3101-1.
  48.  
  49.     D.    CE (European Community) - Electrical Safety Standard IEC 1010-1.
  50.  
  51.     E.    CE (European Community) - Electromagnetic Compatibility Directive 89/336/EEC.
  52.  
  53.     F.    ______________________________________.
  54.  
  55. 1.4    SUBMITTALS
  56.  
  57.     A.    Submit under provisions of Section 01300.
  58.  
  59.     B.    Product Data:  Manufacturer's product data, including physical properties and application instructions.
  60.  
  61.     C.    Shop Drawings:  Prepared specifically for this project; show dimensions and interface with other products.
  62. NOTE ** If required, include paragraph below for 115V/60Hz models, only.
  63.  
  64.     D.    Test Results:  Ensure compliance with NSF 49.
  65.  
  66. 1.5    QUALITY ASSURANCE
  67.  
  68.     A.    Maintain testing facility at manufacturer's place of business for performance testing bench-mounted Class II, Type A and A/B3 safety cabinets.
  69. NOTE ** Include paragraph below for 115V/60Hz models, only.
  70.  
  71.     B.    Safety cabinets wired for 115V/60Hz:
  72.         1.    Meet or exceed NSF Standard 49 minimum requirements and bear NSF Mark.
  73.         2.    Pass complete NSF biological challenge testing at downflow and inflow velocities of plus or minus 10 feet per minute (3.048 m/min) from manufacturer's nominal set point.
  74.         3.    Abridged or altered test methodology is unacceptable.
  75.         4.    Meet or exceed UL Standard 3101-1 and CAN/CSA C22.2 No. 1010.1 minimum requirements.
  76. NOTE ** Include first subparagraph below for projects in the United States, the second subparagraph for projects in Canada.
  77.         5.    Carry ETL Testing Laboratories seal.
  78.         6.    Carry ETL-C Testing Laboratories seal.
  79. NOTE ** Include paragraph below for 230V/50Hz models, only.
  80.  
  81.     C.    Safety cabinets wired for 230V/50Hz:
  82.         1.    Conform to CE (European Community) requirements as tested by Inchcape Testing Services (UK) LTD: Electrical Safety Standard IEC 1010-1 and Electromagnetic Compatibility Directive 89/336/EEC.
  83.  
  84.     D.    Make manufacturing facility, testing facility, and quality control procedures available for owner inspection.
  85.  
  86. 1.6    DELIVERY, STORAGE, AND HANDLING
  87.  
  88.     A.    Deliver bench-mounted biological safety cabinets and accessories free of damage.
  89.  
  90.     B.    Store and handle in a manner to prevent damage to biological safety cabinets, work surfaces, accessories, or adjacent work.
  91.  
  92. 1.7    WARRANTY
  93.  
  94.     A.    Warrant against defects in materials and workmanship on biological safety cabinets and accessories; include labor and replacement parts (except HEPA filters and lamps).
  95.  
  96.     B.    Warranty Period:  Two years from date of installation or three years from date of purchase, whichever is sooner.
  97.  
  98.  
  99. PART 2    PRODUCTS
  100.  
  101. 2.1    MANUFACTURERS
  102.  
  103.     A.    Acceptable Manufacturer:  Provide products made by Labconco Corporation, 8811 Prospect, Kansas City, MO 64132.  ASD.  Tel:  (816) 333-8811 or (800) 821-5525.  Fax:  (816) 363-0130.
  104.  
  105.     B.    Requests for substitutions will be considered in accordance with provisions of Section 01600.
  106. NOTE ** Delete paragraph above or below; coordinate with Division 1 requirements.
  107.  
  108.     C.    Substitutions:  Not permitted.
  109.  
  110.     D.    Provide all biological safety cabinets from a single manufacturer.
  111.  
  112. 2.2    MANUFACTURED UNITS
  113.  
  114.     A.    Bench-Mounted Biological Safety Cabinets:
  115. NOTE ** Select size(s) from three paragraphs below or delete all three paragraphs if drawings indicate size(s).
  116.         1.    38-3/16 inches wide, 58-11/16 inches high, 30-1/4 inches deep (970 x 1490 x 768 mm).
  117.         2.    50-3/16 inches wide, 58-11/16 inches high, 30-1/4 inches deep (1275 x 1490 x 768 mm).
  118.         3.    74-3/16 inches wide, 58-11/16 inches high, 30-1/4 inches deep (1885 x 1490 x 768 mm).
  119. NOTE ** Select paragraph(s) below relevant to project.
  120.         4.    115 volts, 60 Hz models.
  121.         5.    230 volts, 50 Hz models.
  122.  
  123. 2.3    MATERIALS
  124.  
  125.     A.    Steel:  Cold-rolled, 18 gage, epoxy coated.
  126.  
  127.     B.    Stainless Steel:  16 gage, Type 304.
  128.  
  129.     C.    Safety Glass:  3/8 inch (9 mm) tempered.
  130.  
  131. 2.4    COMPONENTS
  132.  
  133.     A.    HEPA Filters:  Industry-standard size, 99.99 percent efficient on particles 0.000012 inch (0.3 micrometers) or larger, without requiring modification for installation.
  134.  
  135.     B.    Fluorescent Lighting:  80-150 foot-candles (861-1615 lm/sq. m) on work surface.
  136.  
  137.     C.    Motor:  1/3 horsepower (248.5 W), thermally protected, capacitor run type.
  138.  
  139.     D.    Wiring Harnesses:  Color coded, alphanumerically labeled for identification; removable wire tags shall not be used.
  140.  
  141.     E.    Speed Control:  Solid state, load rated for 25 amps.  
  142.         1.    Hold motor within 1 volt of setting regardless of line voltage fluctuations above motor voltage setting.
  143.         2.    Start cabinet blower at full line voltage for approximately 1 second before controlling motor voltage at preset level.
  144.  
  145.     F.    Electrical Duplexes with Splash Covers.
  146.         1.    Ground fault interrupters for certain models.
  147.  
  148.     G.    Internal Ductwork:  Epoxy coated steel permanently welded in place.
  149.  
  150.     H.    Internal Balancing Damper:  Sliding guillotine type.
  151.  
  152.     I.    Drain Trough (beneath work tray):  Equipped with 3/8 inch (9.5 mm) ball-type drain valve.
  153. NOTE ** Select the following optional components as required for project.
  154.  
  155.     J.    Service Fixtures:  Chrome-plated brass, quarter-turn, on-off style.
  156.  
  157.     K.    Ultraviolet Light:  Preheat style, type G30T8 or equivalent, 30 watt, 254 nanometer germicidal lamp.
  158.  
  159.     L.    Unassembled Base Stand:  Epoxy coated steel; include assembly instructions.
  160.  
  161. 2.5    FABRICATION
  162.  
  163.     A.    Design:  True laminar (uniform) downflow as defined in NSF Standard 49.
  164.         1.    Nominal Downflow and Inflow:
  165. NOTE ** Select one or both of the following subparagraphs as required for project.
  166.             a.    10 inch (250 mm) Work Access Opening:  80 feet per minute, plus or minus 5 feet per minute (24.38 m/min, plus or minus 1.52 m/min).
  167.             b.    8 inch (200 mm) Work Access Opening:  105 feet per minute, plus or minus 5 feet per minute (32.00 m/min, plus or minus 1.52 m/min).
  168.         2.    Supply and Exhaust HEPA Filters:  Secure by spring-loaded clamps in upper cabinet assembly, compensating for gasket compression; removable from cabinet front.
  169.         3.    Major Electronic Components (Speed Control, Ballasts, Starters, Switches, Motor Capacitors, Circuit Breakers):  House in removable module for service or testing.
  170.         4.    Connections to Speed Control Wiring Harness:  Key to prevent cross connections
  171.         5.    Differential Pressure Gage:  Connect to negative pressure duct.
  172.         6.    Motor Mounting System:  Permanent part of motor housing.
  173.         7.    Electrical Duplex Outlets:  Activated from control panel area with their own circuit breaker.
  174.         8.    Power Cord:  8 feet (2.44 m) long, connect to cabinet by suitable keyed connector (IEC 320 AC inlet connector or equivalent).  Permanently attached cords with strain relief connectors shall not be used.
  175.         9.    Motor/Blower, Lighting System, and Electrical Outlets:  Controlled by separate switches with separate circuit breakers.
  176.         10.    Electrical Connection Box:  Mount on rear panel, removable to allow passage through 30-inch (762 mm) wide doorway.
  177.         11.    Sound Level:  67 dbA or less when measured in accordance with NSF Standard 49.
  178.  
  179.     B.    Liner Assembly; Sides, Rear, and Bottom:  Welded construction, no silicone sealant or solder to seal surfaces.
  180.  
  181.     C.    Upper Cabinet Assembly:  Rolled steel without exposed bolts or screws.  Construct so that positive pressure contaminated plenums are surrounded by negative pressure plenums.  Design so that all major service operations can be performed from front of cabinet.
  182.  
  183.     D.    Work Surface and Front Intake Grille:  Single piece welded construction, radiused corners; removable without disconnecting fasteners.
  184.  
  185.     E.    Sash:  Pivoting type, with provisions to lock in operating, loading, or cleaning positions; positioning shall not require cables, counterweights, or springs.  
  186.         1.    Design so that inner and outer surfaces can be cleaned without disassembly.
  187.         2.    Removal not required for routine filter or motor/blower service.
  188.  
  189.     F.    Electrical Duplex Outlets:  Mount on back wall of stainless steel liner, one on each side.
  190.  
  191.     G.    Cabinet shall accommodate up to 4 service fixtures.
  192.  
  193.  
  194. PART 3    EXECUTION
  195.  
  196. 3.1    EXAMINATION
  197.  
  198.     A.    Verify equipment rough-in before proceeding with work.
  199.  
  200.     B.    Coordinate with other trades for proper installation of plumbing and electrical services and for rough opening dimensions required for hood installation.
  201.  
  202. 3.2    INSTALLATION
  203.  
  204.     A.    Install in accordance with manufacturer's instructions; comply with standards required by authorities having jurisdiction.
  205.  
  206.     B.    Install equipment plumb, square, and straight, without distortion; securely anchor.
  207.  
  208.     C.    Schedule installation to ensure that utility connections are achieved in an orderly and expeditious manner.
  209.  
  210. 3.3    FIELD QUALITY CONTROL
  211.  
  212.     A.    Qualified independent certifier shall certify biological safety cabinet before use, performing tests recommended in NSF International Standard No. 49.
  213.  
  214. 3.4    ADJUSTING AND CLEANING
  215.  
  216.     A.    Clean equipment, casework, countertops, and other surfaces as recommended by manufacturer, rendering work in new and unused appearance.
  217.  
  218.     B.    Clean adjacent construction and surfaces soiled in the course of installation of this work.
  219.  
  220.     C.    Touch up minor damaged surfaces caused by installation.  Replace damaged components as directed by Architect.
  221.  
  222. 3.5    PROTECTION
  223.  
  224.     A.    Provide protective measures to prevent equipment and surfaces from exposure to other construction activity.